Design for on-site test and calibration instrument of forward scatter visibility meter |
Ma Kai,Sun Wenbo,Kong Shiyuan |
Jilin Meteorological Observation Support Center,Changchun 130061 |
|
|
Abstract Aiming at the problems existing in the on-site test and calibration of forward scatter visibility meter,this paper puts forward an improvement plan.According to the working principle of visibility instrument and calibration principle of field test for the scene,the on-site test and calibration instrument is researched and developed.The hardware design for high visibility (zero-scattering signal) and low visibility (high scattering signal) are introduced.And through the experimental data,it proved that the improvement method after device developed is not disturbed by external environment,and the experimental process is robust.The experimental data can be stable rapidly.So this is an on-site calibration method with practical application value for forward scatter visibility meter.
|
Received: 04 June 2018
|
|
|
|
|
[1] |
李春亮,曲世来,等.能见度测量技术100问[M].北京:气象出版社,2009.
|
[2] |
王青梅,谢邦力,梅品琛,等.前向散射仪测量能见度的原理和标定方法探讨[J].气象水文海洋仪器,2001,19(04):10-16.
|
[3] |
王缅,刘文清,陆亦怀,等.气溶胶前向散射大气能见度测量系统传递系数的标定及校准方法[J].光学技术,2008,34(03):334-337.
|
[4] |
王启万.前向散射式能见度仪的发射和接收装置设计[J].气象水文海洋仪器,2005,22(02):1-16.
|
|
|
|