The discussion and research of digital barometers error and uncertainty of measurement results |
Dai Yaru1,2,Sha Yizhuo2,Lu Huiguo1,Che Yunfei1,2 |
1.Chengdu University of Information Technology,Chengdu 610225; 2.CMA Meteorological Observation Centre,Beijing 100081 |
|
|
Abstract There are three digital barometers which are taken as research object to test their performance in this paper.Through performance testing of these barometers,the error of results was analyzed,besides,the uncertainty was evaluated and validated.Through analyzing and researching extensive testing data,it shows that digital barometers errors are mainly from linearity and hysterisis,and then the uncertainty of results is mainly from the repeatability of testing data,the error of standard meter and the resolution ratio of tested instrument.The error of this model digital barometer shows that the error of result is less than 0.17 hPa,the linearity error is less than 0.07 hPa,the hysterisis error is less than 0.03 hPa,and then the uncertainty is not exceed 0.06 hPa after the experiments in high temperature,low temperature and room temperature.
|
Received: 27 March 2014
|
|
|
|
|
[1] |
JJF 1059.1—2012 测量不确定度评定与表示[S].
|
[2] |
李建英,贺晓雷.振筒气压仪性能测试[J].气象科技,2004,32(4):291-293.
|
[3] |
JJF 1094—2002 测量仪器特性评定[S].
|
|
|
|