Uncertainty evaluation of mechanical temperature and humidity meter |
Song Yuan, Wu Lina, Diwu Pengpeng |
Gansu Meteorological Information and Technical Equipment Support Center,Lanzhou 730020 |
|
|
Abstract As the measurement error of mechanical temperature and humidity meter is unavoidable,the evaluation of its measurement uncertainty is an important content of measurement verification.According to JJF 1059.1-2012 Evaluation and Expression of Uncertainty in Measurement,in order to ensure the accuracy and reliability of the measurement data,this paper introduces the general methods and detailed steps of uncertainty evaluation in measurement results of mechanical temperature and humidity meter.With precision dew point meter,LF-300X temperature and humidity test chamber and self-calibration platinum resistance digital thermometer as the standard,and taking WS-1 mechanical temperature and humidity meter as check meter,the evaluation method is described in detail through an example,and the specific evaluation value of extended uncertainty is given out.
|
Received: 23 December 2019
|
|
|
|
|
[1] |
金锐,吴云峰,李丹.CAWS600型自动气象站湿度传感器检定的不确定度分析[J].气象水文海洋仪器,2012,29(2):40-42,45.
|
[2] |
张建敏,罗昶,吕文华.气象计量测试指南[M].北京:中国质检出版社,2011.
|
[3] |
李昕娣,罗怀洁.PTB220型压力传感器检定结果不确定度评估[J].气象科技,2014,42(6):983-985.
|
[4] |
蒋娟萍,卢会国,吴丽娜.现行业务湿度传感器测量不确定度研究[J].成都信息工程学院学报,2015,30(1):37-41.
|
[5] |
姜勇.温度传感器测量不确定度研究[J].气象水文海洋仪器,2019,36(4):81-85.
|
[6] |
张国辉,肖文强,范新国.数字压力计不确定度的评定与表示[J].工程与试验,2016,56(S1):16-18.
|
|
|
|